7
Maintenance
6924-en_V1.3_09/20
83
7.2
Calibrating aerosol sensor
Calibration interval
We recommend calibrating the aerosol sensor every week.
Operating temperature
Ensure that the aerosol spectrometer is at operating temperature during cali-
bration. It will be at operating temperature after having remained on for at
least 30 minutes. If the device is calibrated right after it is switched on, cali-
bration will not be accurate.