6
Operation and settings
6924-en_V1.3_09/20
69
CLEAR | result(ALL, OIL, SALT, USER, PRESS&FLOW)
Deleting measurement results or operator input from current display.
Valid values for parameter 1
| result(ALL, OIL, SALT, USER,
PRESS&FLOW)
Meaning
ALL
All measurement results and opera-
tor input.
OIL
Measurement results for oil aerosol.
SALT
Measurement results for saline
aerosol.
USER
Operator’s reply to a question.
PRESS&FLOW
Values for pressure and flow rate of
breath resistance testing.
REPORT | type(SALT, OIL, BR) | type(OPEN, PRINT, OPEN&PRINT)
Function: Creation of a test report as PDF file with editing options.
Valid values for parameter 1
| type(SALT, OIL, BR)
Meaning
SALT
Measurement results for saline
aerosol.
OIL
Measurement results for oil aerosol.
BR
Values for pressure and flow rate of
breath resistance testing.
Valid values for parameter 2
| type(OPEN, PRINT,
OPEN&PRINT)
Meaning
[blank]
PDF is only saved.
OPEN
Opens the PDF file created.
Prints the PDF file created (printing
on the set standard printer).
OPEN&PRINT
Opens and prints the PDF file cre-
ated (printing on the set standard
printer).