42
CAUTION
Insufficient skin contact of the electrodes
Injury due to unexpected product behaviour as a result of uncontrolled activation of the pros
thesis component.
►
The electrodes are to be placed on intact skin only and with as much electrode-skin contact
as possible.
►
Make sure that skin contact of the electrodes continues to be given when carrying heavy
loads.
►
If the product cannot be controlled correctly using muscle signals, turn off the entire pros
thesis and contact your O&P professional.
CAUTION
Using the prosthesis with pointed or sharp objects (e.g. knives in the kitchen)
Injury due to unintentional movements.
►
Use extra caution when using the prosthesis for handling pointed or sharp objects.
5 Scope of delivery
•
1 pc. 8E38=8*, 8E39=8*, 8E43=8* SensorHand Speed
or
•
1 pc. 8E38=9*, 8E39=9*, 8E41=9* MyoHand VariPlus Speed
•
1 pc. Instructions for use (user)
6 Use
6.1 Terminal device on/off
The ON/OFF button is located directly under the palm of the hand.
The ON/OFF button can be operated by pressing on the prosthetic glove.
Area
Function
Back of the hand
ON (see fig. 1)
Thumb
OFF (see fig. 2)
6.2 Quick-disconnect wrist
The quick-disconnect wrist makes switching between the product and the System Electric Greifer
quick and easy.
If the product is equipped with a quick-disconnect wrist, the following options are available:
•
Bring the terminal device into the most favourable gripping position by rotating it (passive
pronation and supination).
•
Easily disconnect the terminal device from the prosthetic socket to replace it with a System
Electric Greifer.
To do so, turn the terminal device once around its own axis (360°), clockwise or anticlock
wise, until a slight resistance is felt. The terminal device can be removed after overcoming this
resistance. (see fig. 4)
Engaging the quick-disconnect wrist
1) Insert the quick-disconnect wrist of the prosthesis into the lamination ring and press it in
firmly.
2) Turn the prosthesis slightly to the left or right.
3) Pressure triggers the locking mechanism so the hand is firmly anchored in the lamination ring.
INFORMATION: After it engages, check that the prosthesis or System Electric Greifer
is firmly attached.
Содержание 8E38 8 Series
Страница 2: ...1 2 3 4 2...
Страница 207: ...207 1 2 10 N 130 N 300 mm s 300 mm s FlexiGrip 10 N 130 N 1 1 2 10 N 130 N 300 mm s 300 mm s FlexiGrip...
Страница 212: ...212 2 2 1 2 2 AutoControl LowInput 2 1 1 1 2 1 1 MyoBock 1 MyoBock 2 2 1 3 3 VarioControl 1 1...
Страница 213: ...213 1 1 2 100 N 1 1 2 100 N 2 2 1 4 4 VarioDual 2 1...
Страница 214: ...214 2 1 2 1 100 N 1 1 2 2 2 2 1 5 5 DigitalControl 2 1 1 1 2 1 1...
Страница 215: ...215 1 MyoBock 2 2 1 6 6 Double Channel Control 1 3 3 1 3 2 225 3 3...
Страница 216: ...216 3 4 3 5 Ottobock 4 4 1 4 2 1 2 4 3...
Страница 217: ...217 Ottobock Ott bock 4 4...
Страница 218: ...218 4 5 HF Bluetooth WLAN 30 cm HF...
Страница 219: ...219 225 4 6 224...
Страница 220: ...220 Ottobock...
Страница 222: ...222 6 3 1 757L20 EnergyPack 757B2 4 6 1 LED LED LED 2 3 LED 4 LED 5 6 3 1 1 LED 50 50 5 6 3 1 2 LED Ottobock Ottobock...
Страница 223: ...223 6 3 2 757L35 757B35 6 3 2 1 1 2 3 6 3 2 2 6 3 2 3 1 2 LED 50 6 3 2 4 1 2 LED 3...
Страница 224: ...224 7 8 1 Ottobock Derma Clean 453H10 1 N 2 9 24 10 10 1 10 2 10 3 CE Otto Bock Healthcare Products GmbH 2011 65...
Страница 226: ...226 12 12 1 YYYY WW NNN YYYY WW NNN...
Страница 227: ...227...