6. SPECIFICATIONS
Display
Backlit, 7 in. color touchscreen, 800 × 480 pixel resolution. Provides access to the
TCP/IP interface; and displays spectrum, stabilizer, bias, live-time, and dead-time readouts. In
addition to the control and data display screens, you may import a maximum of 9
.JPG
-format
image files for timed, cyclical display.
System Gain Settings
!
Coarse Gain
1, 2, 4, 8, 16, 32, 64, 128
!
Fine Gain
0.5 to 1.1
The available range of gain settings supports all types of HPGe detectors. Specifically, the fol-
lowing maximum energy values are achievable using the standard ORTEC preamplifier (maxi-
mum to minimum gain):
14
!
COAX
187 keV–12 MeV
!
GLP/SLP
16.5 keV–1 MeV
Preamplifier
Software-selectable as either resistive or TRP preamplifier.
Amplifier Input Polarity
Software-selectable.
System Conversion Gain
Software-selectable as 16384, 8192, 4096, 2048, 1024, or 512
channels. Advanced models also include 256, 32768, and 65536 settings.
Digital Filter Shaping-Time Constants
!
Rise time
0.8 ìs to 23 ìs in steps of 0.2 ìs.
!
Flattop
0.3 to 2.4 in steps of 0.1 ìs.
Corrections
!
Dead-Time Correction
Extended live-time correction according to Gedcke-Hale
method. Accuracy: area of reference peak changes <±3% from 0 to 50,000 cps.
!
Zero Dead-Time (ZDT) Correction
When ZDT mode is enabled, dead-time losses are
accounted for by taking very short acquisitions and applying a correction in real time —
as the data are coming in — to the number of counts in the spectrum. This technique
allows the distribution of gamma-rays to change while the acquisition is in progress, yet
14
For the legacy LO-AX and IGLET-X detector types, the respective energy ranges are 94 keV–6 MeV and
8 keV–500 keV.
67
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...