932502G / 0618
7. FIRMWARE COMMANDS AND RESPONSES
where
llllllllll
represents the live time as returned by the SHOW_LIVE command,
tttttttttt
represents the true time for the current device as returned by the SHOW_TRUE command,
aaaaa
represents the active device mask as returned by the SHOW_ACTIVE_DEVICES
command, and
hhhhh
represents the hardware status, which is an ASCII representation of a
16 bit decimal number with the following bit definitions:
Bit 0 (LSB):
Bias supply polarity (0=positive, 1=negative)
Bit 1:
Bias supply overload (0=overload, 1=normal)
Bit 2:
High voltage enabled (0=disabled, 1=enabled)
Bit 3:
Unused
Bit 4:
Amplifier PZ’d since initialization (0=normal, 1=needs
PZ’ing)
Bit 5
Optimization since initialization (0=normal, 1=needed)
Bits 6–7:
Unused
Bit 8:
Amplifier automatic PZ (1=Auto PZ in progress, 0=normal)
Bit 9:
Optimization (0=normal, 1=in progress)
Bits 10–14:
Unused
Bit 15 (MSB):
Reserved
SHOW_THRESHOLD_AUTOMATIC
See also SET_THRESHOLD_AUTOMATIC.
Responses:
$IT<CR>
Automatic threshold is enabled.
$IF<CR>
Automatic threshold is disabled.
SHOW_THRESHOLD_FAST
Returns the current fast channel threshold in the form FMAN_ENA xxxxx or FMAN_DISA
xxxxxx where ENA and DISA indicate whether manual mode is enabled and the numeric
value is the current fast channel threshold. When manual mode is disabled the value is live
so it will change periodically.
Responses:
FMAN_ENA 00000000100
Manual mode is enabled with a setting of 100.
FMAN_DISA 00000000105
Manual mode is disabled with a current value of 105.
SHOW_THRESHOLD_NEGATIVE
See also SET_THRESHOLD_NEGATIVE.
Responses:
THR_NEG 0000379791
Negative threshold is 379791.
SHOW_THRESHOLD_POSITIVE
See also SET_THRESHOLD_POSITIVE.
Responses:
115
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...