932502G / 0618
67. SPECIFICATIONS
Maximum System Throughput
>100,000 cps with LFR off, >34,000 cps with LFR on;
depends on shaping parameters.
Pulse Pileup Rejector
Automatically set threshold. Pulse-pair resolution typically 500 ns. (See
the discussion of the Noise Rejection Level control on page 46.)
Automatic Digital Pole Zero Adjustment
Can be set automatically or manually. Remote
diagnostics via Insight Virtual Oscilloscope mode in MAESTRO.
Digital Gated Baseline Restorer
Software-controlled adjustment of the restorer rate (Fast,
Slow, and Auto).
LLD
Digital lower-level discriminator set in channels. Hard cutoff of data in channels below
the LLD setting.
ULD
Digital upper-level discriminator set in channels. Hard cutoff of data in channels above
the ULD setting.
Ratemeter
Monitor the input count rate on the DSPEC 50 display or computer screen.
Battery
Internal battery-backed up memory to maintain settings in the event of a power
interruption.
Presets
Multiple presets can be set on any or all of the following:
!
Real Time
Stops data collection when the real time reaches this value.
!
Live Time
Stops data collection when the live time reaches this value.
!
ROI
Integral
Integral count stops data collection when the sum of counts in all ROI
channels reaches this value.
!
ROI Peak
Peak count stops data acquisition when the count in any ROI channel reaches
this value (maximum value of 2
31
!
1 counts [2×10
9
]).
!
Uncertainty
Sets the required statistical accuracy on a key peak (for example: stop
counting when the activity of
60
Co is known to be better than 5%).
!
Multi-Nuclide MDA
Stops data collection when the value of the minimum detectable
activity (MDA) for up to 20 user-specified MDA nuclides reaches the preset value. The
presets are implemented in hardware so the computer does not have to poll the
DSPEC 50. See Section 4.10, 4.12, 5.1, 5.2 for details on the MDA preset calculation.
69
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...