DSPEC 50
®
and DSPEC 502
®
Digital Gamma-Ray Spectrometer User’s Manual
932502G / 0618
Figure 41. DSPEC 50 High Voltage Tab.
wide as the peak. Now click the appropriate
Enabled
checkbox to turn the stabilizer on. Until
changed in this dialog, the stabilizer will stay enabled even if the power is turned off. When the
stabilizer is enabled, the
Center Chan
nel and
Width
cannot be changed. For more detailed
information on gain and zero stabilization, see Section D.2.
4.6. High Voltage
Figure 41 shows the High Voltage
tab, which allows you to turn the
HV on or off; set and monitor the
voltage; select the HV
Source
and
Shutdown
mode; and indicate the
detector type; and
Polarity
. Note
that if the detector is attached via
the rear-panel
DIM
connector, some
of these options may be disabled or
auto-selected. For example, the
detector polarity is determined by
the SMART-1 or DIM module.
The
Source
is
Internal
for con-
ventional, non-DIM detectors;
DIM-296
for the Model 296, and
DIM/SMART
for all other DIM-based detectors.
NOTE
NaI detectors require the DIM-POSNAI interface and the
DIM/SMART
source
selection.
The shutdown types are
ORTEC
,
TTL
, and
SMART
. The
ORTEC
mode is used for all
ORTEC detectors except SMART-1 (
SMART
) detectors. For non-ORTEC detectors, check
with the manufacturer. The
TTL
mode is used for most non-ORTEC detectors.
Choose the detector
Polarity
(SMART-1 detectors auto-select this setting). Normally, GEM
(p-type) detectors have a positive signal and GMX (n-type) detectors have a negative signal.
To use a
Sodium Iodide Detector
, mark the checkbox. This changes the gain and zero
stabilizers to operate in a faster mode.
Enter the detector high voltage in the
Target
field, click
On
, and monitor the voltage in the
Actual
field. Click the
Off
button to turn off the high voltage. The HV will not turn on if the
detector is sending a remote shutdown or overload signal. The
Overload
indicator means there
52
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...