932502G / 0618
67. SPECIFICATIONS
HIGH VOLTAGE
(one HV module per chassis)
!
Positive 0–5 kV
Rear-panel SHV connector, 500 V–5 kV. Computer controlled. Only
active when the unit is set for positive bias.
!
Negative 0–5 kV
Rear-panel SHV connector, 500 V–5 kV. Computer controlled. Only
active when the unit is set for negative bias.
SHUTDOWN IN
Rear-panel BNC turns off the bias supply voltage when the detector is warm.
Software-selectable ORTEC or TTL mode (SMART-1 detectors auto-select the SMART shut-
down mode). In ORTEC mode, the detector Bias Shutdown must be connected to this input or
the high voltage will not turn on.
CHANGE SAMPLE OUT
Rear-panel BNC connector, TTL compatible. The output signal is
configurable with the Advanced models as described in Section 4.12.
SAMPLE READY IN
Rear-panel BNC connector accepts TTL level signal from sample
changer. The Advanced models have a counter associated with TTL pulses on the Sample
Ready port.
ETHERNET
Standard 10/100 Mbit Ethernet connection. Link and Activity LEDs are inte-
grated into the connector. Use dynamic or static IP addressing.
USB
Emulates a USB connection. ORTEC C
ONNECTIONS
software supports up to 255 USB-
connected devices per computer.
SD
SecureDigital™ (SD) memory card slot for uploading a maximum of 9 optional
.JPG
-format
image files.
RESET
For system maintenance performed in conjunction with the ORTEC Global Service
Center.
6.1.2. Electrical and Mechanical
Dimensions
42.6 cm W × 35.6 cm D × 15.3 cm H (16.8 in. W × 14.0 in. D × 6.0 in. H).
Weight
11 kg (24.3 lbs).
Operating Temperature Range
0
E
C to +50
E
C, including touchscreen.
71
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...