932502G / 0618
7. FIRMWARE COMMANDS AND RESPONSES
SHOW_ZERO_CHANNEL
Reports the center channel for the stabilizer zero peak. See also SET_ZERO_CHANNEL,
SET_ZERO_WIDTH, and SHOW_ZERO_WIDTH.
Responses:
$C00000087<CR>
Zero channel has not been set.
$C00002089<CR>
Zero channel is channel 2 (lowest possible channel).
$C16382107<CR>
Zero channel is channel 16382 (highest possible
channel).
SHOW_ZERO_STABILIZATION
Reports the state of zero peak stabilization. See also ENABLE_ZERO_STABILIZATION
and DISABLE_ZERO_STABILIZATION.
Responses:
$IT<CR>
Zero stabilization is currently enabled.
$IF<CR>
Zero stabilization is currently disabled.
SHOW_ZERO_WIDTH
Reports the current width for the stabilizer zero peak. See also SET_ZERO_WIDTH,
SET_ZERO_CHANNEL, and SHOW_ZERO_CHANNEL.
Responses:
$C00001088<CR>
Zero width has not been set.
$C00003089<CR>
Zero width is 3 channels (lowest possible width).
$C00256100<CR>
Zero width is 256 channels (highest possible width in
Gauss mode).
$C16383108<CR>
Zero width is 16383 channels (highest possible width
in point mode with zero channel set to 8192).
START
[seg-mask]
Starts the acquisition of spectral data. The optional segment mask is provided for
compatibility with other MCBs and may be any value from 0 to 65535 but is ignored by the
DSPEC 50.
Execution Warnings:
%000004073<CR>
No segment selected (occurs with other warnings).
%000005074<CR>
The acquisition is already started (no changes made).
%000006075<CR>
A preset was exceeded (acquisition was not started).
%000008073<CR>
Unit not optimized since initialization.
%000016076<CR>
Amplifier not PZ’d since initialization.
%000032074<CR>
High voltage is disabled.
The actual response record may be a combination of any of the above records depending on
the warning conditions. For example:
%000048081<CR>
Amplifier not PZ’d and high voltage disabled.
119
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...