APPENDIX C. LIST MODE IN THE DSPEC 50
clock rolls over a RT Word and LT Word are created. The 30-bit real time stamp on the RT word
corresponds to the number of times the 200 ns clock rolls over. The 10 ms real time clock and
the 10 ms live time clocks are cleared when the CLEAR command is sent to the spectrometer.
Otherwise they increment every 10 ms while data is collecting. Using the RT Word as a time
standard and the 200 ns time stamps from the ADC word you can determine the time that any
ADC event occurred, ±200 ns, relative to the beginning of the acquisition.
LT Words are created by the spectrometer at the same time as are RT Words. The 30-bit live
time stamp represents the ADC live seconds since the beginning of the acquisition (or since the
last CLEAR command). By calculating the relative change in live time between any two LT
Words in a list mode stream you can calculate the live time for a given block of data, ±10 ms.
The ADC CRM word contains the value of the spectrometer’s count-rate meter and is generated
at the same time as the RT and LT words (i.e., it always appears next to these words in the list
mode data stream). The count-rate meter is incremented each time the ADC fast channel detects
a pulse. Since many pulses may be rejected before one is converted into an ADC event, this
number is not the simple sum of the ADC events over the past 10 ms; instead it represents the
true input count rate, as well as it can be measured by the instrument. Note that this value is
zeroed after an ADC CRM word is created, therefore it represents only 10 ms of ADC input
counts.
The Counter words contain the value of the external input pulse counters and are generated at
the same time as the RT and LT words. The external pulse counters count the positive pulses at
the external input of the spectrometer. Note that the external inputs must stay high for at least
40 ns for the pulse to be counted. These counters are zeroed after an Ext Counter word is
created, therefore, it represents only 10 ms of external pulses.
Optionally, the external counters can be configured to be a simple level sensitive input. In this
case the counter value is either 0 or 1 depending on the level of the input at the time the counter
word is generated.
137
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...