932502G / 0618
4, 5. MCB PROPERTIES IN MAESTRO
Figure 38. DSPEC 50 ADC Tab.
Figure 39. DSPEC 50A ADC Tab.
10)
At this point, the Resolution Enhancer is now “trained” for the current peak shape para-
meters and the learning mode should be turned off by returning to the Amplifier PRO tab
and unmarking the
Resolution Enhancer
checkbox. The table of adjustments will be
stored in the DSPEC 50's memory.
11)
If you change any parameters that affect peak shape, you must repeat this “training”
procedure.
12)
Once the Resolution Enhancer has been trained and its checkbox has been unmarked, the
Stabilizer tab once again operates on the gain stabilizer (that is, it no longer adds values
to the table of adjustments).
NOTE
The peak selected for the gain stabilizer can be different from the peak used for
training the Resolution Enhancer.
13)
To turn off the Resolution Enhancer, mark its checkbox to turn on the learning mode, go
to the Stabilizer tab and click the
Initialize
button for the gain stabilizer. This will set the
adjustment to zero. Now return to the Amplifier PRO tab and unmark the
Resolution
Enhancer
box.
4.4. ADC
This tab (Fig. 38 and 39 for the Standard and Advanced tabs respectively) contains the
Gate
,
ZDT Mode
,
Conversion Gain
,
Lower Level Disc
riminator, and
Upper Level Disc
riminator
controls. In addition, the current real time, live time, and count rate are monitored at the bottom
of the dialog.
49
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...