Digital Gamma-Ray Spectrometer User’s Manual
932502G / 0618
1.6.4.1. The NORM_CORR Diagnostic Mode
Why is there a
NORM_CORR
mode, and why should you avoid using it? This mode simul-
taneously collects the ZDT spectrum and the conventional uncorrected spectrum. It is useful for
demonstrating that the counts in the uncorrected spectrum divided by the live time is the same
counting rate as the counts in the ZDT spectrum divided by the real time, in the special case of
constant counting rate. Because the error spectrum is not collected in
NORM_CORR
mode, the
standard deviation in the ZDT counts cannot be calculated if the counting rate is varying. Gam-
maVision and ISOTOPIC provide some protection for users if the
ZDT-LTC
mode is inadver-
tently selected. In this case, ISOTOPIC ignores the ZDT spectrum and presumes you intended to
use the uncorrected spectrum in a constant-counting-rate application.
1.6.4.2. To Summarize
Use the ZDT
Off
mode when the counting rate is expected to be constant during the time
taken to acquire the spectrum.
Use the ZDT
CORR_ERR
mode when the counting rate is expected to change or might
change significantly during the time required to acquire the spectrum.
Avoid using the
NORM_CORR
mode because ISOTOPIC v4 will default to analyzing
the LTC spectrum and will ignore the ZDT spectrum.
More Information
For more detailed information, contact your ORTEC representative or go to our website at
www.ortec-online.com
.
!
Application note AN56, “
Loss Free Counting with Uncertainty Analysis Using ORTEC’s
Innovative Zero Dead Time Technique.
”
!
General gamma spectroscopy technical papers in our online publication
Library
.
1.7. Host Computer and Software Requirements
The DSPEC 50 operates on any computer running under Microsoft
®
Windows
®
10, 8, 7, or XP
Professional SP3 when used with C
ONNECTIONS
v9 or higher, and an ORTEC spectroscopy
application such as MAESTRO or GammaVision. The DSPEC 50A or 502A may also be
controlled through the Web interface, which is not Windows as described in Section 5.
14
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...