932502G / 0618
4, 5. MCB PROPERTIES IN MAESTRO
Figure 45. DSPEC 50 Presets Tab.
sets at one time. To disable a preset, enter a value of zero. If you disable all of the presets, data
acquisition will continue until manually stopped.
When more than one preset is enabled (set to a non-zero value), the first condition met during
the acquisition causes the MCB to stop. This can be useful when you are analyzing samples of
widely varying activity and do not know the general activity before counting. For example, the
Live Time
preset can be set so that sufficient counts can be obtained for proper calculation of
the activity in the sample with the least activity. But if the sample contains a large amount of
this or another nuclide, the dead time could be high, resulting in a long counting time for the
sample. If you set the
ROI Peak
preset in addition to the
Live Time
preset, the low-level
samples will be counted to the desired fixed live time while the very active samples will be
counted for the ROI peak count. In this circumstance, the
ROI Peak
preset can be viewed as a
“safety valve.”
The values of all presets for the
currently selected MCB are shown
on the Status Sidebar. These values
do not change as new values are
entered on the Presets tab; the
changes take place only when you
Close
the Properties dialog.
Enter the
Real Time
and
Live
Time
presets in units of seconds
and fractions of a second. These
values are stored internally with a
resolution of 20 milliseconds (ms)
since the MCB clock increments
by 20 ms.
Real time
means elapsed
time or clock time.
Live time
refers
to the amount of time that the MCB is available to accept another pulse (i.e., is not busy), and
is equal to the real time minus the
dead time
(the time the MCB is not available).
Enter the
ROI Peak
count preset value in counts. With this preset condition, the MCB stops
counting when any ROI channel reaches this value unless there are no ROIs marked in the
MCB, in which case that MCB continues counting until the count is manually stopped.
Enter the
ROI Integral
preset value in counts. With this preset condition, the MCB stops
counting when the sum of all counts in all ROI channels (regardless of the number of ROIs)
reaches this value. This has no function if no ROIs are marked in the MCB.
55
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...