DSPEC 50
®
and DSPEC 502
®
Digital Gamma-Ray Spectrometer User’s Manual
932502G / 0618
Figure 47. Nuclide Report Tab.
presets specific to that geometry, efficiency calibration, and nuclide library.
You can then recall this tailored analysis options file as needed.
4.11. Nuclide Report Tab
Figure 47 shows the Nuclide
Report tab. The Nuclide Report
displays the activity of up to nine
(9) user-selected peaks. Once the
report is set up, the two lowest-
energy ROIs and their respective
activity readouts are displayed
on the DSPEC 50's Spectrum
screen.
The peak area calculations in the
hardware use the same methods
as the MAESTRO
Peak Info
option (see the MAESTRO user
manual), so the Nuclide Report
display is the same as the
Peak
Info
display on the selected peak in the spectra stored in the computer. The calculated value is
computed by multiplying the net peak count rate by a user-defined constant. If the constant
includes the efficiency and branching ratio, the displayed value is the activity. You enter the
nuclide label and the activity units. The report format and calculations are discussed in
Appendix D.
IMPORTANT
The entries you make on this screen are saved in the MCB memory, and
are
not dynamically calculated.
If you change the energy calibration (i.e., if the
peak locations shift), the Nuclide Report may no longer be valid.
4.11.1. Add New
You can add Nuclide Peaks to the report manually or by selecting the peaks from the current
working library. The spectrum must be energy calibrated to use the library method.
4.11.1.1. Defining Peaks Manually
To manually define peaks, enter the
Nuclide
name, ROI
Low
(start) and
High
(end) channels,
multiplicative
Factor
and
Units
in the Report section; then click
Add New
. All nuclides in the
table use the same units, so that value need only be entered once.
58
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...