DSPEC 50
®
and DSPEC 502
®
Digital Gamma-Ray Spectrometer User’s Manual
932502G / 0618
of values are presented:
OK
or
ERR
, and numeric value. The SOH parameters return either
OK
or
ERR
. If the state is
OK
, the parameter stayed within the set limits during the spectrum
acquisition. If the parameter varied from the nominal value by more than the allowed limit, the
ERR
is set until cleared by the program. The numeric values are displayed in the units reported
by the DSPEC 50. The
Security
,
Detector temperature
, and
Live detector temperature
are used by SMART-1 detectors and display
N/A
for non-SMART-1 detectors. However, in
this release of the DSPEC 50, the
Detector temperature
and
Live detector temperature
return
N/A
for both SMART-1 and non-SMART-1 detectors.
Detector State of Health
Returns OK or an error message describing a problem with detector power or bias.
Counter A (Sample Ready) (Advanced Models Only)
Displays the current input counter value for the Sample Ready port. This counter is cleared
and started with spectrum acquisition.
Counter B (Gate) (Advanced Models Only)
Displays the current input counter value for the Gate port. This counter is cleared and started
with spectrum acquisition.
+24 volts
This is the current value of the +24 volt supply.
+12 volts
This is the current value of the +12 volt supply.
!
12 volts
This is the current value of the
!
12 volt supply.
!
24 volts
This is the current value of the
!
24 volt supply.
High Voltage
This is the current value of the high voltage bias supply.
4.9. Presets
Figure 45 shows the Presets tab. MDA presets are shown on a separate tab.
The presets can only be set on an MCB that is not acquiring data (during acquisition the preset
field backgrounds are gray indicating that they are inactive). You can use any or all of the pre-
54
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...