DSPEC 50
®
and DSPEC 502
®
Digital Gamma-Ray Spectrometer User’s Manual
932502G / 0618
NetCounts
is computed with the following equation:
GrossCounts
is the sum of the counts in the ROI, excluding the first and last 3 channels of
the ROI.
Background
is:
ROIWidth
is:
Uncertainty
(in percent) is calculated as follows:
Peak
is the position of the maximum count and is computed with the following equation:
MaximumROIChan
is the channel in the ROI with the most counts. If there are no data, the
center channel of the ROI is used.
EnergySlope
and
EnergyIntercept
are the energy calibration values as entered on the
digiDART-LF keypad or by software. If the values are not present, the result is given in
channels.
132
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...