DSPEC 50
®
and DSPEC 502
®
Digital Gamma-Ray Spectrometer User’s Manual
932502G / 0618
Note that the above responses can be combined to indicate a combination of warnings such as:
%000048081<CR>
Amplifier was not pole zeroed nor was high voltage enabled. The
START was attempted anyway.
%000037079<CR>
Device already started and high voltage not enabled. The
command was ignored.
%000053077<CR>
Device already started, high voltage not enabled and amplifier
was not pole zeroed. The command was ignored.
%000038080<CR>
Device preset already exceeded and high voltage not enabled. The
command was ignored.
%000054078<CR>
Device preset already exceeded, high voltage not enabled and
amplifier was not pole zeroed. The command was ignored.
Power-Up Alert:
Power-up just occurred and the selftest results are:
%001000070<CR>
All power-up selftest passed.
%003000072<CR>
Battery backed-up data lost.
Command Syntax Errors:
%129001082<CR>
Invalid command verb.
%129002083<CR>
Invalid command noun.
%129003084<CR>
Invalid command verb and noun.
%129004085<CR>
Invalid command modifier.
%129005086<CR>
Invalid command verb and modifier.
%129006087<CR>
Invalid command noun and modifier.
%129007088<CR>
Invalid command verb, noun, and modifier.
%129132087<CR>
Invalid command (verb, noun, and modifier valid but not
together).
Communication Errors:
%130128084<CR>
Command checksum incorrect (only when optional checksum
provided).
Execution Errors:
%131128085<CR>
Invalid 1st command parameter.
%131129086<CR>
Invalid 2nd command parameter.
%131130078<CR>
Invalid 3rd command parameter.
%131132080<CR>
Invalid number of command parameters.
%131134082<CR>
Invalid device or segment selected.
%131135083<CR>
Command not allowed while acquisition in progress.
%131136084<CR>
Command not allowed in current mode of operation.
%131137085<CR>
Hardware failure detected while processing command.
80
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...