DSPEC 50
®
and DSPEC 502
®
Digital Gamma-Ray Spectrometer User’s Manual
932502G / 0618
Example:
SET_SHUT 0
Sets the SHUTDOWN to TTL.
SET_SHUT 1
Sets the SHUTDOWN to ORTEC.
SET_SHUT 2
Sets the SHUTDOWN to SMART.
SET_SHUTDOWN_ORTEC
This sets the bias shutdown mode to ORTEC.
SET_SHUTDOWN_SMART
Sets the bias shutdown mode to SMART.
SET_SHUTDOWN_TTL
This sets the bias shutdown mode to TTL.
SET_THRESHOLD_FMAN
value
Sets the manual fast channel threshold to the specified value. The value is in machine units
and should be set relative to the automatic fast channel threshold value.
SET_THRESHOLD_NEGATIVE
value
Sets the negative threshold to value. The negative threshold is normally set automatically by
the DSPEC 50. See also ENABLE_ THRESHOLD_AUTOMATIC and SHOW_
THRESHOLD_NEGATIVE.
SET_THRESHOLD_POSITIVE
value
Sets the positive threshold to value. The positive threshold is normally set automatically by
the DSPEC 50. See also ENABLE_THRESHOLD_AUTOMATIC and SHOW_
THRESHOLD_POSITIVE.
SET_TIME
hour,min,sec
Sets the time stored in the battery backed-up system clock to the specified values. hour can
be any value from 0–23; min and sec can be any value from 0–59. The current date and time
are stored for a device when an acquisition is started. See also SHOW_TIME, SET_DATE,
SHOW_DATE, SET_TIME_START, and SHOW_TIME_START.
Non-USB Execution Errors:
%131137085<CR>
The time could not be set due to a hardware malfunction.
Hardware service might be required.
98
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...