DSPEC 50
®
and DSPEC 502
®
Digital Gamma-Ray Spectrometer User’s Manual
932502G / 0618
The
Uncertainty
preset stops acquisition when the statistical or counting uncertainty of a user-
selected net peak reaches the value you have entered. Enter the
Preset in %
value as percent
uncertainty at 1 sigma of the net peak area. The range is from 99% to 0.1% in 0.1% steps. You
have complete control over the selected peak region. The region must be at least 7 channels
wide with 3 channels of background on each side of the peak. Note that MAESTRO calculates
this preset once per 40 seconds. Therefore, the software will continue data acquisition up to
40 seconds after the preset has been reached, and the uncertainty achieved for a high count-rate
sample may be lower than the preset value.
Use the
Start Chan
nel and
Width
fields to enter the channel limits directly, or click
Suggest
Region
. If the marker is positioned in an ROI around the peak of interest,
Suggest Region
reads the limits of the ROI with the marker and display those limits in the
Start Chan
and
Width
fields. The ROI can be cleared after the preset is entered without affecting the
uncertainty calculation. If the marker is not in positioned in an ROI: for calibrated spectra, the
start channel is 1.5 times the FWHM below the marker channel, and the width is 3 times the
FWHM; for uncalibrated spectra, the region is centered on the peak located within two
channels of the marker and as wide as the peak.
The net peak area and statistical uncertainty are calculated in the same manner as for the
MAESTRO
Peak Info
command.
4.10. MDA Preset
The MDA preset (Fig. 46, 48, 50) can monitor up to 20 nuclides at one time, and stops data
collection when the values of the minimum detectable activity (MDA) for
all
of the user-
specified MDA nuclides reach the needed value. Presets are expressed in Bq, and are evaluated
every 40 seconds. The detector must be calibrated for energy in all spectroscopy applications,
and for efficiency in all applications but MAESTRO.
The MDA presets are implemented in the MCB (i.e., the entries you make on this screen are
saved in the MCB memory), and have no direct link to MDA methods selected in the analysis
options for applications such as GammaVision, ScintiVision, ISOTOPIC, etc. The MDA preset
calculation uses the following formula:
where:
a
,
b
, and
c
are determined by the MDA criteria you choose.
56
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...