3. INSTALLATION
Figure 30. Change MCB Number or Description.
3.5.2. Customizing ID Numbers and Descriptions
If you wish, you can change the instrument ID Numbers and Descriptions by double-clicking
an instrument entry in the Configure Instruments dialog. This will open the Change
Description or ID dialog (Fig. 30). It shows the physical MCB location (read-only), and allows
you to change the
ID
Number and
Description
.
Make the desired changes and click
Close
. Any changes you have made to an ID number or
description will then be written back to the corresponding MCB.
If a modified description has already been applied to a particular instrument, you can restore
the default description by deleting the entry in the
Description
field and re-running MCB Con-
figuration. After MCB Configuration runs, the default description will be displayed.
3.5.3. Connecting to and Disconnecting from the Computer
The DSPEC 50 can be connected to and disconnected from the computer without shutting
down either. If you disconnect from the computer during data acquisition (leaving the
DSPEC 50 under power so that the high voltage stays on), the DSPEC 50 will continue data
NOTE FOR MULTIPLE USERS ON A NETWORK
There are two ways to reduce the chance that other users will renumber your MCBs:
!
Add the
-I
flag to their MCB Configuration command line, as described in the C
ONNECTIONS
Driver Update Kit instructions. This will allow you to assign whatever ID Numbers you wish,
regardless of the numbers assigned by other users on your network. (Ideally, everyone using
ORTEC instruments on your network should make this change.)
!
To prevent others from renumbering your MCBs (or performing any other actions except read-
only viewing), password-lock your MCBs with the MAESTRO
Lock/Unlock Detector
command. If you lock a detector that will be controlled by a JOB stream, remember to include
the proper password-unlock commands in your
.JOB
file (see the MAESTRO user manual).
37
Содержание DSPEC 50
Страница 4: ... Intentionally blank iv ...
Страница 8: ... Intentionally blank viii ...
Страница 10: ...x ...
Страница 26: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 16 ...
Страница 72: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 62 ...
Страница 76: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 66 ...
Страница 86: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 76 ...
Страница 132: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 122 ...
Страница 136: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 126 ...
Страница 144: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 134 ...
Страница 148: ...DSPEC 50 and DSPEC 502 Digital Gamma Ray Spectrometer User s Manual 932502G 0618 Intentionally blank 138 ...