Chapter 4 Assembly
89
Attach the condenser to the condenser mount.
You can attach the TI-C condenser turret (system condenser) or ELWD-S condenser to the condenser mount.
Additionally, attach the TI-C-LWD condenser lens or the MC TMD2 ELWD condenser lens to the system
condenser.
■
Preparation for attachment
Loosen the condenser clamp screw on the right side
of the condenser holder using a hexagonal
screwdriver.
* The condenser clamp screw is located inside the
hole on the right side of the condenser holder. If
the condenser mount is shifted from the reference
position, the screw will not be visible in the hole.
In that case, loosen the condenser mount rotation
clamp screw, align the positioning groove on the
mount with the positioning pin on the condenser
holder by turning the mount, and then tighten the
condenser mount rotation clamp screw.
■
Attaching the TI-C system condenser
1 Turn the condenser turret to the “A” position
(dummy hole cassette position for bright-field
microscopy).
2 With the “A” label facing the front (towards the
user), insert the dovetail on the system condenser
into the bottom of the condenser holder, and
secure it by tightening the condenser clamp screw.
Attach the condenser turret by sliding it in from the
front.
3 Insert the condenser cassette into the condenser
turret, and secure it with two hexagonal socket
head bolts.
Up to five condenser cassettes can be attached to
the turret. Attach the cassettes so that their
numbering is increased by turning the turret
clockwise (as viewed from above).
4 Screw in the condenser lens into the bottom of the
turret.
Positioning pin
Positioning
groove
Condenser mount
rotation clamp screw
Condenser mount
Condenser
holder
Condenser
clamp screw*
(within side hole)
Condenser
cassette
Condenser lens
Condenser
clamp screw
Dovetail
System condenser
“A” position