Chapter 3 Operation Details
56
3.22
Using the HG Precentered Fiber Illuminator (C-HGFI/C-HGFIE)
Figure 3.22-1 HG Precentered Fiber Illuminator
The specified light source LV-LH50PC may not
provide sufficient brightness to perform epi-fl
microscopy. In such case, use an HG precentered
fiber illuminator that uses a high intensity mercury
lamp as a light source.
3.22.1
Procedure for turning on the power switch
Figure 3.22-2 Procedure for turning on
the power switch
1 Be sure that the fiber illuminator is properly
connected to the MA200 main body.
2 Set the POWER switch to the “|” side to turn on
the fiber illuminator and wait until it starts up.
When the POWER indicator on the fiber
illuminator lights up, the LAMP indicator starts
blinking. Once the illuminator starts up, the LAMP
indicator stops blinking and lights up.
If the C-HGFIE motorized fiber illuminator is in
use, the communication function on the illuminator
does not activate until it starts. Be sure to wait until
the illuminator starts up.
3 Set the POWER switch to the “|” side to turn on
the power on the MA200 main body.
The power LED on the front of the MA200 main
body lights up.
4 To turn off the fiber illuminator, set the power
switch to the “
c
” side in the sequence of the fiber
illuminator, the MA200 main body.
* To use the HG precentered fiber illuminator
correctly, carefully read the instruction manual
supplied with the fiber illuminator and make sure
to follow the instruction.
HG Precentered
Fiber Illuminator
LAMP indicator
POWER indicator
2
Power switch ON
1
Power switch ON