Chapter 4 Assembly
85
4.13
Polarizer/analyzer Unit and Fluorescent Unit
1 Loosen the front-cover fixing bolt with the supplied
hexagonal screwdriver to remove it. Turn to loosen
the BD field changeover lever counter-clockwise,
then remove the lever and front-cover from the
main body.
2 Insert the unit into the slot inside the microscope.
Push the unit into the second click-stop position,
the polarizer/analyzer or fl filter enters the optical
path. Pull out the unit until it hits the click-stop
position, the polarizer/analyzer or the fl filter can
be removed from the optical path.
3 Remove the dust protection cover from the front
cover to hide the operation port, then return the
front cover to the original place to fix it with the
bolt.
To perform the epi-fl microscopy using the
C-HGFI/C-HGFIE HG Precentered Fiber
Illuminator as the external light source, screw the
compensation filter supplied with the fiber
adapter into the fluorescent unit. (Refer to
4.15.1.)
4.13.1 MA2-
λ
P
λ
plate
The MA2-
λ
P
λ
plate can be attached to the
MA2-PA/MA2-UPA polarizer/analyzer unit. Insert the
MA2-
λ
P slider from the front as illustrated.
Polarizer/analyzer unit
Fluorescent unit
λ
plate