The weight unit for the CAL weight may differ from the unit in the device. Conversion
is automatic.
Note:
Error messages, calibration see Chapter
7.12.8
Calibrating with mV/V
The scale can be calibrated without weights. During input of the load cell mV/V value, the
acceleration of gravity at the place of installation can be taken into account.
The STAR load cell data is based on the acceleration of gravity in Hamburg, Germany:
9.81379 m/s
2
.
5.
Press
to select the weight unit.
6. Confirm entries with
or
.
Weight value, weight unit and measuring signal in mV/V corresponding to this
value are displayed in the line [Calibrated at].
The verification is displayed by "Setting Span by load…".
1.
Calculating SPAN value for Max and, if necessary, for the dead see Chapter
2.
Press the [by mV/V] softkey.
3.
Entering the SPAN value for Max and, if necessary, for the subsequent correction of
dead load (see Chapter
Transmitter Series PR 5220