If applicable, press the [No] softkey to keep the original values.
3.
Enter and confirm e.g. the value for 4 mA measured by the connected PLC under
[Measured].
The window for the 2nd value opens.
4. Enter and confirm e.g. the value for 20 mA measured by the connected PLC under
[Measured].
A prompt window opens.
5.
Press the [Yes] softkey to save the settings.
The following window opens:
9 Extended functions
Transmitter Series PR 5220
EN-171
Minebea Intec