16.6 Error numbers @ "LAST_ERROR"
16.6.1
Weighing point error
Number
Display
Cause
2
in use
Weighing point is in use, e. g., by an application.
6
test active
Test is active, no weights.
7
cali active
Calibration is active, no weights.
8
no standstill
No standstill of the scale.
13
tare is active
Tare is active.
15
weight is dimmed
Weight is not legal-for-trade (<0 or >max.).
16
weight has error
Weight error.
17
scale not ready
Scale is not ready.
18
cannot tare below zero
Taring below zero is not possible.
102
timeout
Only on xBPI scales: Time limit exceeded when sending a com-
mand to the scale.
142
calibration active
During calibration, taring and zeroing is not possible.
147
no zeroset
Zeroset outside of the zero setting range is not possible.
149
Busy
The scale is currently busy with another query.
255
hardware error
Weighing point is faulty.
16.6.2
Error in the "EasyFill" application
Number
Display
Cause
0
No error.
1
fatal error
Weight error; weighing point is faulty.
2
Material ID invalid
Number <0 or >10 entered.
3
Material name inva-
lid
Material name is invalid.
4
Set point invalid
Set point + gross value > scale end value, (B1 mode) or gross value.
5
Preset point invalid
Value > set point.
6
OVS invalid
Value > set point.
8
Neg. tol. invalid
Value > set point.
9
Pos. tol. invalid
Value > set point.
10
Sequence number
invalid
Sequence number is invalid.
13
Invalid fieldbus com-
mand
Fieldbus action is invalid (e. g., simultaneous start and stop com-
mand).
15
Cannot read from
earom
Error when reading a material entry from the hard drive memo-
ry (EAROM) → hardware error
Transmitter Series PR 5220
16 Error messages
Minebea Intec
EN-224