VNC text
Error and possible cause
Remote dis-
play
Scale not ready
The scale is not ready for weighing:
-
The instrument is in the warm-up phase.
-
The instrument is in automatic taring mode.
-
The instrument was switched on with the scale loaded.
Switch the device off and on again.
Error <
Wrong serial number
Serial number of scale does not match the number set in the de-
vice.
Error?
16.4 Error messages for Pendeo® load cells
VNC text
Error and possible cause
Remote dis-
play
Overload
The measuring signal is higher than Max + (x d):
-
Wrong setting.
-
Too much weight on the scale.
Error 2
Value exceeds display
The weight value is not displayed:
-
Too many digits have been set.
Error 4
Incompatible units
Incompatible units:
-
Incorrect calibration values e.g. due to incorrect calibration
Error 8
No values from scale
No communication with Pendeo load cell:
-
Cable break.
-
Internal scale error.
-
The junction box is not connected to the supply voltage.
Error 9
Scale not ready
The scale is not ready for weighing:
-
At least 1 load cell gives an error status or is defective (no
communication).
Error <
Wrong configuration
The number of load cells does not match the configuration.
Error =
Wrong serial number
Serial number of scale does not match the number set in the de-
vice.
Error?
16 Error messages
Transmitter Series PR 5220
EN-221
Minebea Intec