7.12.12
Calculating the test value
The calculation of the test value is called by pressing the [CalcTest] softkey.
The maximum load (Max) is displayed with the ID
without a weight unit. The value
determined during calibration by pressing the [CalcTest] softkey after starting the test is
displayed.
Depending on the settings under
-[Weighing point]- [Calib]- [Param]- [Test mode]
the following is displayed by calling the Test with the
key later on:
-
with [Absolute] the maximum load
-
with [Relative] the deviation from the test value
7.12.13
Saving the calibration
Quit calibration by pressing the
softkey.
A message is displayed, indicating that the value for Max cannot be changed as
long as linearization is active.
7 Commissioning
Transmitter Series PR 5220
EN-89
Minebea Intec