Note:
When setting [IP Address], [Subnetmask], and [Default gateway], please consult with
your system administrator.
Press
to return to the Setup menu and to save the changes.
7.15.5
Configuring limit values
Note:
This menu item is only available if under
-[Operating parameter]- [Application]
"Standard" has been selected.
The parameters for limits are set under this menu item.
Each limit consists of a switch-on and a switch-off point for definition of a hysteresis. The
3 pairs of values must be entered according to the same principle. The limit values always
refer to the gross weight. For the SPM addresses for the limits, see Chapter
NOTICE
The limit values of an xBPI weighing point are scale-specific.
The scale must be active when entering the limit values.
The scale and the unit must not be changed after configuration.
The following settings are required: [Weighingpoint/xBPI-Scale] - [Setup] -
[Configuration] - [Application settings] - [Number of units] "1 Weight"
Define the parameters for limits under
-[Limit parameter].
For the configuration the following order must be followed:
1. Define limits.
2. Define an action.
3. Determine a condition.
4. Save parameters.
7 Commissioning
Transmitter Series PR 5220
EN-137
Minebea Intec
Содержание PR 5220 Series
Страница 26: ...3 6 Overview of connections 3 Device description Transmitter Series PR 5220 EN 23 Minebea Intec ...
Страница 234: ......