7.15.5.1
Defining limits
Example 1:
The output signal (Limit 1 out) of limit 1 switches OFF above a weight (Wgt) of 900 kg.
The output signal (Limit 2 out) of Limit 2 switches OFF below 290 kg.
The two limit values have a hysteresis of 10 kg.
In the event of a power failure both outputs turn to "off" ("OFF"), thus indicating underfill
and overfill simultaneously.
Example 2:
If the Limits 1 and 2 are the same for "On" and "Off" (on = off),
-
switches output 1 (Limit 1 out) ON if the weight (Wgt) exceeds the value.
-
switches output 2 (Limit 2 out) OFF if the weight falls below the value.
Transmitter Series PR 5220
7 Commissioning
Minebea Intec
EN-138