NOTICE
Flow rate measurement problems may occur.
It is imperative to connect the equipotential bonding conductor between PR 1626/
60 and mounting rail.
4.4.7.9
Connecting an analog platform (series CAP…)
You can connect an analog platform to the device.
NOTICE
The cable colors shown above are valid for a CAPP4 500 x 400 and a
CAPP1 320 x 420.
The assignments of cable colors are listed in the load platform operating
instructions.
Connection designations
Combics 1 3-devices
PR 5220
BR_POS
V+ LC SUPPLY
SENSE_POS
S+ SENSE
OUT_POS
M+ SIGN.
OUT_NEG
M- SIGN.
SENSE_NEG
S- SENSE
BR_NEG
V- LC SUPPLY
The cable screens must be connected to the grounding terminal of the device. If the
measuring lines (+M, -M) are screened individually, these screens must be connected to
the GND terminal as well (see also Chapter
).
4 Device installation
Transmitter Series PR 5220
EN-45
Minebea Intec