[Ambient conditions]
This parameter is used to define the ambient conditions of the scale.
Possible Selections: very stable condition, stable condition, unstable condition, very
unstable condition
[W&M]
Setting for legal-for-trade mode.
Note:
The transmitter of the series PR 5220 are not approved for calibration.
[Unbal. Check deviat.]
The plausibility check is activated when the average deviation is >0%. The average
deviation of the individual load cells is calculated.
Setting range: 0 100%.
[Standstill time]
The parameters [Standstill time] and [Standstill range]are used to define the stability of
the scale (stable balance).
The input for the parameter [Standstill time] is expressed in seconds.
Valid range: 0.00 2 s
If the tome is set to "0" there is no check. The standstill time must not be less than the
measuring time.
[Standstill range]
As long as the weight fluctuations remain within this range, the device is determined to
be stable.
The input for the parameter [Standstill range] is expressed in "d."
Valid range: 0.01 10.00 dc.
[Tare timeout]
Timeout for a tare/zeroset command that cannot be executed (e.g. due to mechanical
instability of the scale, incorrect filter setting, resolution too high, standstill condition too
strict).
The input is done as seconds.
Valid range: 0.0 <2.5> 25 s.
At 0.0 s taring is only carried out when the scale is already stable.
Transmitter Series PR 5220
7 Commissioning
Minebea Intec
EN-122