17 - 32
Operator’s Manual
17 Probes and Biopsy
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put on the sterile probe sheath.
b.
Hold the probe by one hand, select the proper needle-guided bracket, and hold it with the
other hand. Align the narrow end tab of the needle-guided bracket with the groove of the
probe, then push the needle-guided bracket forward, making the tabs and the grooves of
the needle-guided bracket to match with the grooves and tabs of the probe.
c.
Check manually to confirm that the needle-guided bracket is securely installed on the
probe.
d.
Select a proper guiding block and push it into the groove above the angle block, and
clamp it tightly.
e.
Insert a biopsy needle with the same specification as that of the guiding block into the
hole of the guiding block.
2.
Remove the needle-guided bracket:
a.
Remove the guiding block slightly along the direction of the needle's tail.
b.
Separate the residual part of the needle-guide bracket and the transducer from the needle.
7
Specification of angle block (60°)
The corresponding biopsy angle is 60°.
Содержание Imagyn 7
Страница 2: ......
Страница 14: ...This page intentionally left blank...
Страница 20: ...This page intentionally left blank...
Страница 54: ...This page intentionally left blank...
Страница 72: ...This page intentionally left blank...
Страница 118: ...This page intentionally left blank...
Страница 126: ...This page intentionally left blank...
Страница 196: ...This page intentionally left blank...
Страница 240: ...This page intentionally left blank...
Страница 280: ...This page intentionally left blank...
Страница 298: ...This page intentionally left blank...
Страница 406: ...This page intentionally left blank...
Страница 416: ...This page intentionally left blank...
Страница 491: ......
Страница 492: ...P N 046 019593 01 3 0...