17 Probes and Biopsy
Operator’s Manual
17 - 17
3.
Disinfect the probe by using an appropriate high-level disinfectant or system. For how to use a
high-level disinfectant or system, see the operator’s manual provided by the manufacturer.
Prepare a disinfectant by using sterile distilled or softened water when necessary.
–
Soaking: Immerse the probe head in the disinfectant and shake the probe appropriately to
remove any bubbles on the probe surface. For details about the probe immersion duration,
see the operator’s manual provided by the manufacturer.
Observe the graph here carefully to immerse the probe. Only soak parts of the probe
below the strain relief.
–
Wiping: Use a market disinfection wipe product or sterile disposable lint-free soft cloth
wetted with disinfection spray and wipe all surfaces of the probe for a duration according
to the manufacturer instructions.
4.
Rinse the probe thoroughly by using a large amount of clean water (about 7.5 L/2 gallons) at
room temperature for about 30 s to remove the residual disinfectant. Repeat the operation
twice. Or follow the disinfectant manufacturer’s instructions regarding rinsing.
5.
Dry the probe by wiping with a piece of clean disposable lint-free soft cloth. Do not dry the
probe by heating.
6.
Check whether the probe has defects such as peeling, rifts, bumps, cracks, or liquid spill. If
such defects exist, the probe has reached the end of its service life. In this case, stop using it
and contact the Mindray service department.
7.
Store the probe in a cool, clean and dry environment. And repeat the cleaning and disinfection
process before the next use.
Sterilization of a critical probe
CAUTION
Repeated sterilization will eventually damage the probe, please check the
probe’s performance periodically.
For intra-operative probes, they have to be thoroughly cleaned and sterilized after completing each
examination.
Perform the following procedure:
1.
Wear a pair of gloves to prevent infection.
2.
Clean the probe thoroughly in accordance with the cleaning procedure before sterilization.
3.
Sterilize the probe by using an appropriate sterilant or system.
1
Strain relief
2
Probe handle
3
Fluid level
1
2
3
Содержание Imagyn 7
Страница 2: ......
Страница 14: ...This page intentionally left blank...
Страница 20: ...This page intentionally left blank...
Страница 54: ...This page intentionally left blank...
Страница 72: ...This page intentionally left blank...
Страница 118: ...This page intentionally left blank...
Страница 126: ...This page intentionally left blank...
Страница 196: ...This page intentionally left blank...
Страница 240: ...This page intentionally left blank...
Страница 280: ...This page intentionally left blank...
Страница 298: ...This page intentionally left blank...
Страница 406: ...This page intentionally left blank...
Страница 416: ...This page intentionally left blank...
Страница 491: ......
Страница 492: ...P N 046 019593 01 3 0...