17 - 68
Operator’s Manual
17 Probes and Biopsy
c.
Hold the probe. Press the biopsy needle pressure position to separate it from the V-shaped
guiding block. Put the needle into the V-shaped guiding block of the needle guided-
bracket.
d.
Hold the probe, and release the pressure position of the needle. Adjust the needle-shaped
adjusting nut manually (following the direction of the arrow). The needle moves smoothly
at the vertical direction due to its gravity.
2.
Remove the needle-guided bracket:
a.
Hold the probe. Press the biopsy needle pressure position to separate the biopsy needle
from the pressure position. Separate the bracket and the probe from the needle.