Operator’s Manual
1 - 1
1
Safety Precautions
1.1
Meaning of Safety Symbols
1.2
Safety Precautions
Please observe the following precautions to ensure patient and operator’s safety when using this
system.
DANGER
Do not operate this system and probes in an atmosphere containing flammable
gases or liquids such as anesthetic gases, hydrogen, and ethanol, because
there is danger of explosion.
WARNING
•
Do connect the power plug of this system to wall receptacles that meet the
ratings indicated on the rating nameplate. If adapters or multifunctional
receptacles are used, it may cause the leakage current to exceed the
safety requirement.
•
In the environment that patient is 1.5 meters around, connect peripherals to
the auxiliary power outlet, or power the peripherals by auxiliary output
cable or isolation transformer complied with IEC60601-1 or the power input
of the same safety level.
Symbol
Description
Type-BF applied part
The ultrasound probes connected to this system are type-BF applied parts.
The PCG leads within this system is type-BF applied part.
The ECG leads within this system is type-BF applied part.
Caution!
General warning sign.
Patient injury or tissue damage from ultrasonic radiation. The ALARA
principle must be practiced when operating the ultrasound system.
Содержание Imagyn 7
Страница 2: ......
Страница 14: ...This page intentionally left blank...
Страница 20: ...This page intentionally left blank...
Страница 54: ...This page intentionally left blank...
Страница 72: ...This page intentionally left blank...
Страница 118: ...This page intentionally left blank...
Страница 126: ...This page intentionally left blank...
Страница 196: ...This page intentionally left blank...
Страница 240: ...This page intentionally left blank...
Страница 280: ...This page intentionally left blank...
Страница 298: ...This page intentionally left blank...
Страница 406: ...This page intentionally left blank...
Страница 416: ...This page intentionally left blank...
Страница 491: ......
Страница 492: ...P N 046 019593 01 3 0...