17 Probes and Biopsy
Operator’s Manual
17 - 45
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put on the sterile probe sheath.
b.
Select a proper needle-guided bracket, and match the locating groove with the tab of the
transducer. Mount the bracket onto the transducer.
c.
Tighten the pinch nut of the needle-guided bracket to confirm that the needle-guided
bracket is properly installed on the transducer.
d.
Adjust the dial scale to the required needle type shift.
e.
Adjust the needle angle to the proper shift as required (loosen the nut first, and then
tighten the nut based on the shift you need).