12 - 2
Operator’s Manual
12 Smart Pelvic Floor (Smart Pelvic)
12.2 3D/4D image data
VR Mode
Perform the following procedure:
1.
Select probe and GYN/pelvic floor exam mode.
2.
Enter 3D/4D preparation status, and then acquire 3D/4D image to enter Smart Pelvic.
a.
Auto: Select [Pelvic] under the [Smart Scene 3D] scenario, and then press <Update>. The
system enters [Smart Pelvic] tab automatically.
b.
Manual: select a non-pelvic subpreset under the [Smart Scene 3D] scenario, press
<Update>, and tap [Smart Pelvic] tab.
3.
On image-frozen mode, adjust image parameters.
On 4D imaging mode, if necessary, tap [Timer] to turn on the timer, and freeze the image as
needed.
4.
Tap [Auto Measure] to start automatic measurement.
5.
If necessary, tap [Edit], points are uniformly distributed on VR image automatically. Roll the
trackball to drag the points. Tap [Edit] again to exit. After editing, tap [Auto Measure] to
update the measurement result.
During the editing, tap [Undo] to undo last operation, and tap [Undo All] to undo all
operations.
6.
Tap [Rest]/[Valsalva]/[Contraction] to mark current image status.
Tap [Hide] to hide measurement results if necessary.
CMPR Mode
Perform the following procedure:
1.
Tap [Multi-plane Mode] on smart pelvic VR mode to enter CMPR mode.
The reference line with an arrow and the other two sampling lines generated by the system are
displayed on the screen. With the reference line being the start line, the three lines correspond
to Open/Closing/Closed images in turn.
2.
If necessary, perform the following operations.
–
Tap [Ref. Image] to switch between MPRs, and tap [Open]/[Closing]/[Closed] to switch
activated window.
–
Tap [Set Ref. Lines] to set the reference line.
–
Tap to display the planes in 3*3 mode.
–
Rotate the knob under [Line Extension] on the touch screen to adjust the line length.
–
Rotate the knob under [Rotate RL] on the touch screen to adjust the line angle.
–
Rotate the knob under [Spacing] on the touch screen to adjust the spacing between the
three lines.
–
Rotate the knob under [Thickness] on the touch screen to adjust the imaging thickness.
–
Rotate the <4D> knob to scroll CMPR image.
–
Tap [Auto Measure] to start automatic measurement.
3.
Tap [Back] to go back to VR mode.
Parameter Settings
•
VOI Thickness
Rotate the knob under [VOI Thickness] to adjust the thickness of VOI.
Содержание Imagyn 7
Страница 2: ......
Страница 14: ...This page intentionally left blank...
Страница 20: ...This page intentionally left blank...
Страница 54: ...This page intentionally left blank...
Страница 72: ...This page intentionally left blank...
Страница 118: ...This page intentionally left blank...
Страница 126: ...This page intentionally left blank...
Страница 196: ...This page intentionally left blank...
Страница 240: ...This page intentionally left blank...
Страница 280: ...This page intentionally left blank...
Страница 298: ...This page intentionally left blank...
Страница 406: ...This page intentionally left blank...
Страница 416: ...This page intentionally left blank...
Страница 491: ......
Страница 492: ...P N 046 019593 01 3 0...