17 Probes and Biopsy
Operator’s Manual
17 - 61
NGB-040
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put on the sterile probe sheath.
b.
Hold a needle-guided bracket, and match the tab with the groove of the transducer. Mount
the bracket onto the transducer.
c.
Rotate the clamping nuts of the guided bracket on the right and left side to fix the bracket
and the transducer. Rotate the needle-type adjusting nut to the ultimate position as shown
in the figure.
1
Left/right clamp
2
Left/right clamp pinch nut
3
Guiding bulges
4
Needle clamp cover
5
Needle type adjusting nut
6
Separation pressure position
7
Guiding hole
8
V-shaped guiding hole
9
Angle shift sign
10
Angle pinch nut
11
Angle adjusting base
12
Angle adjusting block
<2>
<3>
<4>
<5>
<6>
<7>
<8>
<9>
<10>
<11>
<12>
<1>
Содержание Imagyn 7
Страница 2: ......
Страница 14: ...This page intentionally left blank...
Страница 20: ...This page intentionally left blank...
Страница 54: ...This page intentionally left blank...
Страница 72: ...This page intentionally left blank...
Страница 118: ...This page intentionally left blank...
Страница 126: ...This page intentionally left blank...
Страница 196: ...This page intentionally left blank...
Страница 240: ...This page intentionally left blank...
Страница 280: ...This page intentionally left blank...
Страница 298: ...This page intentionally left blank...
Страница 406: ...This page intentionally left blank...
Страница 416: ...This page intentionally left blank...
Страница 491: ......
Страница 492: ...P N 046 019593 01 3 0...