17 - 62
Operator’s Manual
17 Probes and Biopsy
d.
Hold the transducer. Press the biopsy needle pressure position to separate it from needle
guided V-shaped block.
e.
Put the needle into the needle guided-bracket, and the needle leans to V-shaped block.
f.
Hold the transducer, and release the pressure position of the needle. Adjust the needle-
type adjusting nut manually (following the direction of the arrow). The needle moves
smoothly at the vertical direction due to its gravity.
2.
Remove the needle-guided bracket:
a.
Hold the transducer, and press the pressure position of the needle to separate the biopsy
needle from the biopsy needle pressure position.