5-80
MICRONIX USA, LLC
www.micronixusa.com
Rev: 3.02
Microsteps
During Motion
Real-time
Program
Global
Set
Read
Set
Read
Set
Read
Set
Read
Command
Description:
This command sets the current Microstepping value. Max
determined by UMX. 4096 at full current.
Returns:
A read operation returns the microstepping value
for the specified
axis.
Syntax:
nUSTx
–
Standard syntax
nUST?
–
Read microstepping rate
0USTx
–
Missing axis number, all axes set to x microstepping
Error [#]:
UST
?
–
Read operation with missing axis number [27]
nUST
–
Missing microstep parameter [28]
Parameter
Description:
n[int]
–
Axis number
x[float]
–
Microsteps
?
–
Read microstep value
Parameter
Range:
n
–
0 to 99
x
–
1 to UMX (4096)
Related Commands: UMX
Example:
1UST2000
|Axis 1, Set 2000 microsteps [degrees/s]
-
5UST?
|Axis 5, Read microstep value