Replacing the Chamber Cap O-Ring
Fine fibers and particles between the O-ring and its sealing surfaces can
cause leaks, as can scratches or cuts in the O-ring or in the metal surfaces.
When it is necessary to replace the O-ring, follow these steps.
1.
Remove the O-ring from its groove by pushing a sharp tool into the ring
and prying it out.
Be careful not to scratch the metal surface of the chamber cap.
Scratches could result in an imperfect seal.
Figure 5-2. Removing the Chamber Cap O-Ring
AccuPyc 1330
Maintenance
September 1996
5-15