Transmission port
waiting for Xon
Cause:
The receiving device stopped transmission by sending an
Xoff, and hasn’t resumed the transmission by sending an
Xon.
Action:
None; when the receiving device is ready for more data,
it should send the pycnometer an Xon.
The following message is displayed on the printer or on the screen for trans-
mitted data.
No collected data to report, or all runs excluded
Cause:
This message is printed or transmitted if you requested a
report but there is either no data available or data has
been excluded via review mode.
Action:
Initiate an automatic operation or return to review mode
and include at least one run.
AccuPyc 1330
Error and Status Messages
September 1996
5-13
Содержание AccuPyc 1330
Страница 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Страница 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Страница 61: ...CHAPTER 5 TROUBLESHOOTING AND MAINTENANCE Troubleshooting Error and Status Messages Maintenance ...
Страница 83: ...CHAPTER 6 ORDERING INFORMATION ...
Страница 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Страница 92: ...APPENDIX B ANALYSIS THEORY ...
Страница 96: ...APPENDIX C CALIBRATION THEORY ...
Страница 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Страница 108: ...APPENDIX E PRINTER CHARACTER SET ...
Страница 111: ...INDEX ...