
Display
Entry
Enter the equilibration rate and press
.
Range:
0.0001 to 9.0000 psig/min
0.0007 to 62.05 kPag/min
A high rate will produce faster results, but
results may not be as precise as desired.
The lowest rates may cause errors when
some materials (such as materials with
appreciable vapor pressures, closed cell
foams, or organics) are analyzed.
The run precision feature enables early
termination of the analysis when certain
criteria are met. When you select run
precision, the analysis will be terminated
after five consecutive runs are within the
specified tolerance.
If you select run precision, you should
request a large number (50 to 99) of runs.
If you select a small number of runs, the
analysis will stop when the number you
entered is reached even though the specified
tolerance has not been met.
Select yes or no, then press
.
This prompt applies for analysis only.
Equilibration rate:
0.005 psig/min
ENTER
Use run precision?
No
ENTER
Editing the Setup Parameters
AccuPyc 1330
4-20
September 1996
Содержание AccuPyc 1330
Страница 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Страница 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Страница 61: ...CHAPTER 5 TROUBLESHOOTING AND MAINTENANCE Troubleshooting Error and Status Messages Maintenance ...
Страница 83: ...CHAPTER 6 ORDERING INFORMATION ...
Страница 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Страница 92: ...APPENDIX B ANALYSIS THEORY ...
Страница 96: ...APPENDIX C CALIBRATION THEORY ...
Страница 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Страница 108: ...APPENDIX E PRINTER CHARACTER SET ...
Страница 111: ...INDEX ...