CHANGING THE ANALYSIS MODULE
Analysis modules, which house the cell chamber, are available in various sizes,
enabling you to analyze different size samples with precision. To change the
analysis module:
1.
Turn off the gas supply at the regulator mounted to the gas cylinder by turn-
ing the Cylinder Shut-Off Valve fully clockwise.
Figure A-1. Turning Off Gas Supply
2.
Press
on the pycnometer keypad to enter manual mode. Then
press VENT to open the vent valve, EXPAND to open the expansion valve,
and FILL to open the fill valve.
The power cord should be disconnected from the unit before disconnect-
ing the gas supply. Failure to do so could result in electrical shock.
3.
Observe the pressure reading. When the pressure drops to less than 1 psig,
place the power switch on the back of the pycnometer in the off (O) posi-
tion. Remove the power cord.
1
Inlet
Vent
Cylinder Shut-Off
Valve
AccuPyc 1330
APPENDIX A
September 1996
A-1