TROUBLESHOOTING AND MAINTENANCE
This chapter describes common operational problems and their solutions, er-
ror and status messages, and maintenance procedures. If further assistance is
needed after following the procedures in this chapter, contact a Micromeritics
Service Representative.
Troubleshooting
Operating problems encountered with the pycnometer are usually easily cor-
rected. Typical problems and the steps required to correct them are described
in the following table.
Table 5-1. Common Operational Problems
What Happened
Why
What To Do
Nothing happens when
power switch is turned
on (ie: no lights on
schematic display, noth-
ing on LCD screen).
Power cord not fully
inserted at one end or
the other.
Insert power plug
firmly into outlet
socket; insert unit
connector firmly into
power connector
opening.
No power at outlet.
Plug in lamp or small
appliance to test outlet.
If there is no power,
contact electrician.
Plug prongs bent so
that contact not made
at outlet.
Wiggle power plug at
outlet while watching
display. If display
comes on, have
electrician adjust
prongs or replace
outlet or plug.
Power cord damaged.
Have electrician check
cord using test meter.
Replace if defective.
AccuPyc 1330
Troubleshooting
September 1996
5-1
Содержание AccuPyc 1330
Страница 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Страница 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Страница 61: ...CHAPTER 5 TROUBLESHOOTING AND MAINTENANCE Troubleshooting Error and Status Messages Maintenance ...
Страница 83: ...CHAPTER 6 ORDERING INFORMATION ...
Страница 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Страница 92: ...APPENDIX B ANALYSIS THEORY ...
Страница 96: ...APPENDIX C CALIBRATION THEORY ...
Страница 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Страница 108: ...APPENDIX E PRINTER CHARACTER SET ...
Страница 111: ...INDEX ...