USR_ERR: No data to review
Cause:
You tried to review data for an automatic operation when
there were no data to review.
Action:
Abandon request.
USR_ERR: Number of runs must be = 5
Cause:
You tried to enable run precision without increasing the
number of runs to at least five.
Action:
Increase the number of runs to five or abandon request.
USR_ERR: Out of range
Cause:
You tried to enter a value that is out of the valid range.
Action:
Enter a value in the specified range (refer to Chapter 4).
USR_ERR: Pressure overrange
Cause:
A pressure overrange occurred but was left uncorrected.
Action:
Return to manual mode and vent the system.
ZEROERR: Eq failure
Cause:
The system failed to equilibrate within 1000 seconds.
Action:
Check the pycnometer for leaks. Make sure the pycnome-
ter has been placed in a draft-free environment.
ZEROERR: Overrange
Cause:
A pressure overrange occurred while zeroing.
Action:
Check for a fill valve leak. If there is none, contact a
Micromeritics Service Representative.
ZEROERR: Underrange
Cause:
An error occurred in the pressure measurement electronics
while zeroing, causing a pressure underrange to occur.
Action:
Call a Micromeritics Service Representative.
March 2000
5-11
AccuPyc 1330
Error and Status Messages
Содержание AccuPyc 1330
Страница 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Страница 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Страница 61: ...CHAPTER 5 TROUBLESHOOTING AND MAINTENANCE Troubleshooting Error and Status Messages Maintenance ...
Страница 83: ...CHAPTER 6 ORDERING INFORMATION ...
Страница 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Страница 92: ...APPENDIX B ANALYSIS THEORY ...
Страница 96: ...APPENDIX C CALIBRATION THEORY ...
Страница 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Страница 108: ...APPENDIX E PRINTER CHARACTER SET ...
Страница 111: ...INDEX ...