Analysis or Calibration Parameters
Press
to display or edit analysis or calibration parameters.
Display
Entry
Press
until
Analysis parameters is
displayed. Then press
.
Enter the number of purges to be performed
and press
.
Range:
0 to 99
Purging cleans the cell and expansion
chambers before an analysis begins. The
greater the number of purges, the cleaner
the sample will be when analyzed.
Enter the fill pressure and press
.
Range:
0 to 19.850 psig
0 to 136.86 kPag
Generally, the greater the fill pressure, the
easier it is to measure the volume precisely.
However, a lower pressure may be desirable
for some samples.
Enter the number of runs to be performed
and press
.
Range:
1 to 99
Enter the fill pressure and press
.
Range:
0 to 19.850 psig
0 to 136.86 kPag
2
Set-up type?
Analysis parameters
CHOICE
ENTER
Number of purges:
3
ENTER
Purge fill pressure:
19.500 psig
ENTER
Number of runs:
3
ENTER
Run fill pressure:
19.500 psig
ENTER
AccuPyc 1330
Editing the Setup Parameters
September 1996
4-19
Содержание AccuPyc 1330
Страница 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Страница 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Страница 61: ...CHAPTER 5 TROUBLESHOOTING AND MAINTENANCE Troubleshooting Error and Status Messages Maintenance ...
Страница 83: ...CHAPTER 6 ORDERING INFORMATION ...
Страница 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Страница 92: ...APPENDIX B ANALYSIS THEORY ...
Страница 96: ...APPENDIX C CALIBRATION THEORY ...
Страница 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Страница 108: ...APPENDIX E PRINTER CHARACTER SET ...
Страница 111: ...INDEX ...