Reviewing and Editing Calibration Results and Parameters
Press
to review or edit calibration results.
Display
Entry
The volume of the calibration standard is
displayed.
Select either cell chamber or expansion
chamber, then press
.
or
Either cell or expansion volume is
displayed depending on the selection made
at the previous prompt. [N] is replaced with
the number of the run. The deviation from
the mean is also displayed.
You may choose to exclude the displayed
cell or expansion volume from the
calculated average by pressing
.
When a volume is excluded from the
average, an asterisk appears next to it.
Likewise, a cell or expansion volume that
was previously excluded from the calculated
average may be included by pressing
.
When you press
, a new deviation is
calculated and displayed.
Press
to display values for the next
run.
Press
when you wish to exit review
and edit mode and return to display mode.
When you press
, all collected data
are automatically recalculated and all data
reduction messages are added back into the
queue.
5
Cal std volume:
(volume)
Which chamber?
Cell
ENTER
Cell [ N] = (volume)
Dev [ N] = (deviation)
Exp [ N] = (volume)
Dev [ N] = (deviation)
CHOICE
CHOICE
CHOICE
ENTER
SAVE
SAVE
AccuPyc 1330
Reviewing and Editing Data
September 1996
4-9
Содержание AccuPyc 1330
Страница 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Страница 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Страница 61: ...CHAPTER 5 TROUBLESHOOTING AND MAINTENANCE Troubleshooting Error and Status Messages Maintenance ...
Страница 83: ...CHAPTER 6 ORDERING INFORMATION ...
Страница 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Страница 92: ...APPENDIX B ANALYSIS THEORY ...
Страница 96: ...APPENDIX C CALIBRATION THEORY ...
Страница 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Страница 108: ...APPENDIX E PRINTER CHARACTER SET ...
Страница 111: ...INDEX ...