Greasing the Chamber Cap O-Ring
A greasing disk is included in the accessory kit shipped with the pycnometer.
Use the disk to grease the O-ring as described below.
Rubber gloves should be worn to prevent contaminating the grease with
oil from your fingertips. Carefully apply a thin layer of grease; both too
much and too little grease can cause problems. Too much grease may al-
ter cell volume. Too little grease results in an imperfect seal and leaks.
1.
Remove the chamber cap by turning it counterclockwise then lifting off.
2.
Using your fingertip, apply a thin layer of Dow Corning high vacuum
grease (or equivalent) to the greasing disk.
3.
Insert the greasing disk into the chamber cap and turn slightly to apply a
very light coating of grease to the O-ring.
Apply a VERY LIGHT coating of grease to the O-ring. There should be
no visible ridges of grease; excess grease may alter analysis results.
Figure 2-10. Greasing the O-Ring in Chamber Cap
4.
Replace the chamber cap.
Greasing Disk
O-ring
Chamber Cap
Equipment Setup
AccuPyc 1330
2-10
September 1996
Содержание AccuPyc 1330
Страница 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Страница 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Страница 61: ...CHAPTER 5 TROUBLESHOOTING AND MAINTENANCE Troubleshooting Error and Status Messages Maintenance ...
Страница 83: ...CHAPTER 6 ORDERING INFORMATION ...
Страница 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Страница 92: ...APPENDIX B ANALYSIS THEORY ...
Страница 96: ...APPENDIX C CALIBRATION THEORY ...
Страница 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Страница 108: ...APPENDIX E PRINTER CHARACTER SET ...
Страница 111: ...INDEX ...